3 Language Dictionary
124 Keysight Models 6811C, 6812C, and 6813C Programming Guide
FREQuency:MODE
This command determines how the output frequency is controlled during a
triggered output transient. The choices are as follows:
– FIXed: The output frequency is unaffected by a triggered output transient.
– STEP: The output frequency is programmed to the value set by
FREQuency:TRIGgered when a triggered transient occurs.
– PULSe: The output frequency is changed to the value set by
FREQuency:TRIGgered for a duration determined by the pulse commands.
– LIST: The output frequency is controlled by the frequency list when a triggered
transient occurs.
FREQuency:SLEW
This command sets the rate at which frequency changes for all programmed
changes in output frequency. Instantaneous frequency changes can be obtained
by sending MAXimum or INFinity. The SCPI keyword INFinity is represented by the
number 9.9E37.
Command syntax [SOURce:]FREQuency:MODE <mode>
Parameters FIXed | STEP | PULse | LIST
*RST value FIXed
Examples
FREQ:MODE FIX
Query syntax [SOURce:]FREQuency:MODE?
Returned parameters <CRD>
Related commands FREQ FREQ:TRIG
Command syntax [SOURce:]FREQuency:SLEW[:IMMediate] <NRf+> | INFinity
Parameters 0 to 9.9E37 | MAXimum | MINimum | INFinity
*RST value MAXimum
Examples
FREQ:SLEW:IMM 75 FREQ:SLEW MAX
Query syntax [SOURce:]FREQuency:SLEW[:IMMediate]?
Returned parameters <NR3>
Related commands FREQ FREQ:SLEW:MODE