3 Language Dictionary
136 Keysight Models 6811C, 6812C, and 6813C Programming Guide
LIST:CURRent
This command sets the sequence of peak output current list points. The current
points are given in the command parameters, which are separated by commas.
The order in which the points are entered determines the sequence in which they
are output when a list is triggered. Changing list data while a subsystem is in list
mode generates an implied ABORt.
LIST:CURRent:POINts?
This query returns the number of points specified in LIST:CURRent. Note that it
returns only the total number of points, not the point values.
Command syntax [SOURce:]LIST:CURRent <NRf+>{, <NRf+>}
Parameters MINimum | MAXimum |
0 to 3.25 (6811C including all options)
0 to 6.50 (6812C including all options)
0 to 13.0 (6813C all options except Option 019)
0 to 16.0 (6813C Option 019)
Unit A (peak amperes)
*RST value 1
Examples
LIST:CURR 2.5, 3.0, 3.5
LIST:CURR MAX, 3.5, 2.5, MIN
Query syntax [SOURce:]LIST:CURRent?
Returned parameters <NR3>{, <NR3>}
Related commands LIST:CURR:POIN? LIST:COUN LIST:DWEL LIST:STEP
Query syntax [SOURce:]LIST:CURRent:POINts?
Returned parameters <NR1>
Examples
LIST:CURR:POIN?
Related commands LIST:CURR