3 Language Dictionary
164 Keysight Models 6811C, 6812C, and 6813C Programming Guide
VOLTage:SENSe:DETector
VOLTage:ALC:DETector
These commands select the type of closed-loop feedback (voltage control loop)
that is used by the output power circuits of the ac source. The commands are
interchangeable; they both perform the same function. The following closed-loop
feedbacks can be selected:
– RTIMe: This feeds the instantaneous output voltage back to the error amplifier
and compares it to the reference waveform.
– RMS: This converts the rms output voltage to DC and compares it to a DC
reference.
Command syntax [SOURce:]VOLTage:SENSe:DETector <type>
[SOURce:]VOLTage:ALC:DETector <type>
Parameters RTIMe | RMS
*RST value RTIMe
Examples
VOLT:SENS:DET RTIM VOLT:ALC:DET RMS
Query syntax [SOURce:]VOLTage:SENSe:DETector?
[SOURce:]VOLTage:ALC:DETector?
Returned parameters <CRD>
Related commands VOLT:SENS:SOUR