OUTPut:PROTection:CLEar
This command clears the latch that disables the output when an overvoltage (OV),
overcurrent (OC), overtemperature (OT), remote inhibit (RI), or power rail fault
condition is detected. All conditions that generated the fault must be removed
before the latch can be cleared. The output is then restored to the state it was in
before the fault condition occurred.
OUTPut:PROTection:DELay
This command sets the delay time between the programming of an output change
that produces a CL or UNREG status condition and the recording of that condition
by the Questionable Status Condition register. The delay prevents momentary
changes in status that can occur during programming from being registered as
events by the status subsystem. In most cases these temporary conditions are not
considered an event, and to record them as such would be a nuisance.
Command syntax OUTPut:PROTection:CLEar
Parameters None
Examples
OUTP:PROT:CLE
Related commands OUTP:PROT:DEL *SAV *RCL
Command syntax OUTPut:PROTection:DELay <NRf+>
Parameters 0 to 100 | MAXimum | MINimum
Unit s (seconds)
*RST value 100 milliseconds
Examples
OUTP:PROT:DEL 75E-1
Query syntax OUTPut:PROTection:DELay?
Returned parameters <NR3>
Related commands OUTP:PROT:CLE *SAV *RCL