3 Language Dictionary
106 Keysight Models 6811C, 6812C, and 6813C Programming Guide
OUTPut:IMPedance
This command enables or disables the output impedance programming capability
of the ac source.
OUTPut:IMPedance:REAL
This command sets the real part of the output impedance of the ac source.
OUTPut:IMPedance:STATe must be enabled for the programmed impedance to
affect the output.
Command syntax OUTPut:IMPedance[:STATe] <bool>
Parameters 0 | 1 | OFF | ON
*RST value OFF
Examples
OUTP:IMP 1 OUTP:IMP ON
Query syntax OUTPut:IMPedance[:STATe]?
Returned parameters 0 | 1
Related commands OUTP:IMP:REAL OUTP:IMP:REAC
Command syntax OUTPut:IMPedance:REAL <NRf>
Parameters 0 to 1
Unit Ω (ohms)
*RST value 0
Examples
OUTP:IMP:REAL 0.25
Query syntax OUTPut:IMPedance:REAL?
Returned parameters <NR3>
Related commands OUTP:IMP OUTP:IMP:REAC