EasyManuals Logo

Keysight x-series User Reference Guide

Keysight x-series
1138 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #1107 background imageLoading...
Page #1107 background image
12 Disturbance Analyzer Measurement
Save
If the save is initiated via SCPI, and the file already exists, the file and the directory will be
overwritten.
Using the C: drive is strongly discouraged, since it runs the risk of being overwritten during an
instrument software upgrade.
Both single and double quotes are supported for any filename parameter over SCPI.
Initial S/W Revision A.14.00
Edit Header
This key is only available in the Disturbance Analyzer measurement. It enables you to edit the
measurement results header information.
Key Path
Save, Export Data,
Initial S/W Revision A.14.00
Title
Provides a title of Meas Result file with a maximum length up to 100 characters. The descriptions will be
stored in the exported Meas Result file as the title and can be displayed in the active function area by
selecting them as the active function.
Key Path
Save, Export Data, Edit Header
Remote Command
:MMEMory:RESults:TITLe “string”
:MMEMory:RESults:TITLe?
Example :MMEM:RES:TITL “Advanced Company EMC Report”
Preset (null String), not affected by Mode Preset, preset by Restore Mode Defaults.
State Saved Saved in instrument state
Initial S/W Revision A.14.00
Client
Provides the Client information with a maximum length up to 100 characters. The descriptions will be
stored in the exported Meas Result file as the Client field and can be displayed in the active function area
by selecting them as the active function.
Key Path
Save, Export Data, Edit Header
Remote Command
:MMEMory:RESults:CLIent “string”
:MMEMory:RESults:CLIent?
Example :MMEM:RES:CLI XYZ Corp Limited”
Preset (null String), not affected by Mode Preset, preset by Restore Mode Defaults.
State Saved Saved in instrument state
Initial S/W Revision A.14.00
EMI Receiver Mode Reference 1107

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals