8 Frequency Scan Measurement
View/ Display
Flat Monochrome
Selects a format that is like Flat Color. But only black is used (no colors, not even gray), and no fill.
Key Path
Save, Screen Image, Themes
Example MMEM:STOR:SCR:THEM FMON
Readback Flat Mono
Initial S/W Revision Prior to A.02.00
Backlight
Accesses the display backlight on/off keys. This setting may interact with settings under the Windows
"Power" menu.
When the backlight is off, pressing ESC, TAB, SPACE, ENTER, UP, DOWN, LEFT, RIGHT, DEL, BKSP, CTRL,
or ALT turns the backlight on without affecting the application. Pressing any other key will turn backlight on
and could potentially perform the action as well.
Key Path
View/Display, Display, System Display Settings
Remote Command
:DISPlay:BACKlight ON|OFF
:DISPlay:BACKlight?
Preset ON (Set by Restore Misc Defaults)
Initial S/W Revision Prior to A.02.00
Backlight Intensity
An active function used to set the backlight intensity. It goes from 0 to 100 where 100 is full on and 0 is off.
This value is independent of the values set under the Backlight on/off key.
Key Path
View/Display, Display, System Display Settings
Remote Command
:DISPlay:BACKlight:INTensity <integer>
:DISPlay:BACKlight:INTensity?
Example DISP:BACK:INT 50
Preset 100 (Set by Restore Misc Defaults)
Min 0
Max 100
Initial S/W Revision Prior to A.02.00
Meters Display
Enables you to turn on/off the meters display. When ON and no scan or final measurement is running, the
meters will take up the right side of the display. When OFF, the frequency scan expands to fill the entire
EMI Receiver Mode Reference 603