5 Mode Functions
Mode Setup
Max 100 s
Default Unit s
Backwards
Compatibility SCPI
:SENSe1|SENSe2|[SENSe]:SWEep:TIME
Backwards Compatibility
SCPI Notes
This command above is included for ESU compatibility. The suffix 1|2 is irrelevant to the
measurement, they are added solely for remote language compatibility.
Initial S/W Revision A.07.00
Peak Hold Time
Accesses a menu that lets you set the peak hold time of the meters. If set to Adjustable, the max hold line
and max result of the meters will be reset for each due of a peak hold time. If set to infinite, the max hold
line and max result would not be reset.
Key Path
Mode Setup, Meters Control
Remote Command
[:SENSe]:METer:PHOLd:TYPE INFinite|ADJustable
[:SENSe]:METer:PHOLd:TYPE?
Example MET:PHOL:TYPE INF
MET:PHOL:TYPE?
Preset INFinite
State Saved Saved in instrument state
Range Adjustable|Infinite
Initial S/W Revision A.07.00
Adjustable
Sets the Peak Hold Time to Adjustable.
Key Path
Mode Setup, Meters Control, Peak Hold Time
Example MET:PHOL:TYPE ADJ
Readback Text <time>
Initial S/W Revision A.07.00
Key Path
Mode Setup, Meters Control, Peak Hold Time
Remote Command
[:SENSe]:METer:PHOLd:ADJustable <time>
[:SENSe]:METer:PHOLd:ADJustable?
Example MET:PHOL:ADJ 500 ms
MET:PHOL:ADJ?
EMI Receiver Mode Reference 253