EasyManuals Logo

Keysight x-series User Reference Guide

Keysight x-series
1138 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #554 background imageLoading...
Page #554 background image
8 Frequency Scan Measurement
Save
.lim file format
This is a legacy format which allows files saved from older analyzers to be loaded into the X-Series.
Design of files in this format is not recommended.
Signal List
Pressing this key selects Signal List as the data type to be exported with a save request. The next step is to
select the Save As key in the Save Data menu.
See "Signal List File" on page 554 for an example of data contained in the list.
Key Path
Save, Data
Remote Command
:MMEMory:STORe:SLISt <filename>
Example MMEM:STOR:SLIS C:\My Documents\EMI\data\mySignalList2.csv
exports the Signal List to the file mySignalList2.csv in the default path.
Notes The default path is at “My Documents\EMI\data\FSCAN\SignalList\SignalL_xxxx.csv
If the save is initiated via SCPI, and the file already exists, the file will be overwritten.
Using the C: drive is strongly discouraged, since it runs the risk of being overwritten during an
instrument software upgrade.
Both single and double quotes are supported for any filename parameter over SCPI.
Initial S/W Revision A.07.00
Signal List File
A Signal List file contains a copy of one of the signal lists obtained during measurement. The first four lines
are system required header lines, and must be in the correct order.
SignalList DataFileTypeName
EMI:FSCAN Mode Name: Measurement Name
A.01.05N9020A InstrumentVersion ModelNumber
K03 EMC SEC BBA P13 P08 BAB 5031 OptionList FileFormatVersion
The data follows as comma separated values. A keyword DATA precedes the data. Here is an example of
signal list file with five signals on it:
SignalList
EMI:FSCAN
A.01.05,N9020A
B1X P13 EA3 BBA B2X ERC UK6 BAB EMC K03 B25 503 P08 526 508 513 S40 P26 B1C HBA P03 PFR ,01
Detector Limit,1,2,3
Final Detector,Peak,Sample,Average
Y Axis Unit,DecibelMicrovolt
DATA
554 EMI Receiver Mode Reference

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals