8 Frequency Scan Measurement
Sweep/ Control
Couplings When stopped, the label on the key changes to Start.
Backwards
Compatibility SCPI
:ABORt
Backwards Compatibility
SCPI Notes
This SCPI is remote language compatible with ESU.
Initial S/W Revision A.07.00
Clear List and Start
This key will perform the following when pressed:
1. Prompt you for confirmation to delete all signals in signal list with the following message:
2. After confirmation, performs the selected scan sequence. This will turn the label on the Start key to
Stop. If the selected Scan Sequence is Scan, and the scan is set to continuous, you need to stop the
measurement with the Stop key. For all other cases, the scan sequence will stop when the measurement
has completed.
Key Path
Sweep/Control
Remote Command
INITiate:FSCan:CLEar:IMMediate
Example INIT:FSC:CLE:IMM
Dependencies
This key is disabled in two conditions:
1.A Scan sequence is running. If the grayed out key is pressed, an advisory message is generated. If
the equivalent SCPI command is sent, this same message is generated as part of a “–221, Settings
conflict” warning.
2.A Scan Sequence is set to Remeasure. If the grayed out key is pressed, an advisory message is
generated. If the equivalent SCPI command is sent, this same message is generated as part of a “–
221, Settings conflict” warning.
Initial S/W Revision A.07.00
Single (Meters)
Performs a single measurement at meters.
Key Path
Sweep/Control
EMI Receiver Mode Reference 575