EasyManuals Logo

Keysight x-series User Reference Guide

Keysight x-series
1138 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #376 background imageLoading...
Page #376 background image
7 Trigger Functions
Trigger
:TRIGger[:SEQuence]:ATRigger:STATe?
Example TRIG:ATR:STAT ON
TRIG:ATR 100 ms
Notes The "time that the analyzer will wait" starts when the analyzer is ready for a trigger, which may be
hundreds of ms after the data acquisition for a sweep is done. The "time" ends when the trigger
condition is satisfied, not when the delay ends.
Preset Off, 100 ms
State Saved Saved in instrument state
Min 1 ms
Max 100 s
Default Unit s
Initial S/W Revision Prior to A.02.00
Trig Holdoff
Sets the holdoff time between triggers. When the trigger condition is satisfied, the trigger occurs, the delay
begins, and the holdoff time begins. New trigger conditions will be ignored until the holdoff time expires.
For a free-running trigger, the holdoff value is the minimum time between triggers.
Key Path
Trigger, Auto/Holdoff
Remote Command
:TRIGger[:SEQuence]:HOLDoff <time>
:TRIGger[:SEQuence]:HOLDoff?
:TRIGger[:SEQuence]:HOLDoff:STATe OFF|ON|0|1
:TRIGger[:SEQuence]:HOLDoff:STATe?
Example TRIG:HOLD:STAT ON
TRIG:HOLD 100 ms
Dependencies Unavailable if the selected Input is BBIQ. If this is the case, the key is grayed out if it is pressed the
informational message Feature not supported for this Input” is displayed. If the SCPI command is
sent, the error Settings conflict; Feature not supported for this Input” is generated.
Preset Off, 100 ms
State Saved Saved in instrument state
Min 0 s
Max 0.5 s
Default Unit s
Initial S/W Revision Prior to A.02.00
376 EMI Receiver Mode Reference

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)