5 Mode Functions
Mode Setup
SENS:FSC:SLIS:COUP:MET?
Couplings Couple Meters to Marker key will be turned off when this key is turned on.
Preset ON
State Saved Saved in instrument state.
Range On|Off
Initial S/W Revision A.07.00
Couple Meters to Marker
If this is turned on, the measurement will automatically couple Meters frequency to frequency of the
selected marker.
Key Path
Marker
Remote Command
:CALCulate:FSCan:MARKer:COUPle:METer ON|OFF|1|0
:CALCulate:FSCan:MARKer:COUPle:METer?
Example CALC:FSC:MARK:COUP:MET 1
CALC:FSC:MARK:COUP:MET?
Couplings Couple Meters to Signal List key will be turned off when this key is turned on.
Preset OFF
State Saved Saved in instrument state.
Range On|Off
Backwards
Compatibility SCPI
:CALCulate[1]|2:MARKer:COUPled[:STATe] ON|OFF|1|0
:CALCulate[1]|2:MARKer:COUPled[:STATe]?
Initial S/W Revision A.13.00
Autorange Atten 10 dB Minimum
Toggles the autorange attenuation protection to have at least 10 dB always set on the input.
Key Path
Mode Setup
Remote Command
[:SENSe]:METer:ATTenuation:PROTection ON|OFF|1|0
[:SENSe]:METer:ATTenuation:PROTection?
Example MET:ATT:PROT ON
MET:ATT:PROT?
Couplings When ON, an attenuation of at least 10 dB is always set on the input when autorange is running
during meters measurement, prescan and the final measurement.
Preset OFF
State Saved Saved in instrument state
Backwards
:INPut[1]|2:ATTenuation:PROTection
256 EMI Receiver Mode Reference

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals