3 Programming the Analyzer
STATus Subsystem
Example STAT:QUES:CAL:SKIP:COND?
Preset 0
Status Bits/OPC
dependencies
Sequential command
Initial S/W Revision Prior to A.02.00
Questionable Calibration Skipped Enable
This command determines which bits in the Questionable Calibration Skipped Condition Register will set
bits in the Questionable Calibration Skipped Event register, which also sets bit 11 of the Questionable
Calibration Register. The variable <integer> is the sum of the decimal values of the bits you want to enable.
Mode All
Remote Command
:STATus:QUEStionable:CALibration:SKIPped:ENABle <integer>
:STATus:QUEStionable:CALibration:SKIPped:ENABle?
Example STAT:QUES:CAL:SKIP:ENAB 1 Can be used to query if an EMI alignment skipped condition is
detected
Preset 32767
Min 0
Max 32767
Status Bits/OPC
dependencies
Sequential command
Initial S/W Revision Prior to A.02.00
Questionable Calibration Skipped Event Query
This query returns the decimal value of the sum of the bits in the Questionable Calibration Event register.
The register requires that the associated PTR or NTR filters be set before a condition register bit can set a
bit in the event register. The data in this register is latched until it is queried. Once queried, the register is
cleared.
Mode All
Remote Command
:STATus:QUEStionable:CALibration:SKIPped[:EVENt]?
Example STAT:QUES:CAL:SKIP?
Preset 0
Status Bits/OPC
dependencies
Sequential command
Initial S/W Revision Prior to A.02.00
EMI Receiver Mode Reference 119

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals