8 Frequency Scan Measurement
Meas Setup
New avg = ((K–1)Old avg + New data)/K,
where K is the number of averages accumulated. (In continuous mode, once K has reached the # of Scans,
K stays at that value, providing a continuous running average.)
Key Path
Meas Setup
Remote Command
[:SENSe]:FSCan:SWEep:COUNt <integer>
[:SENSe]:FSCan:SWEep:COUNt?
Example FSC:SWE:COUN 10
FSC:SWE:COUN?
Dependencies This key is grayed out when a scan sequence is running. If the grayed out key is pressed, an advisory
message is generated. If the equivalent SCPI command is sent, this same message is generated as
part of a “–221, Settings conflict” warning.
Preset 1
State Saved Saved in instrument state.
Min 1
Max 999
Backwards
Compatibility SCPI
:SENSe1|SENSe2:SWEep:COUNt
[:SENSe]:SWEep:COUNt
Backwards Compatibility
SCPI Notes
This command above is included for ESU compatibility. The suffix 1|2 is irrelevant to the
measurement, they are added solely for remote language compatibility.
Initial S/W Revision A.07.00
Scan Type
Specifies the scan type, either as Smooth, Discrete (Stepped) or Time Domain The scan will go through all
of the range tables that are currently on, and scans the specified range using the range table parameters.
Key Path
Meas Setup
Remote Command
[:SENSe]:FSCan:SCAN:TYPE SMOoth|DISCrete|TDOMain
[:SENSe]:FSCan:SCAN:TYPE?
Example FSC:SCAN:TYPE SMO
FSC:SCAN:TYPE?
Dependencies This key is grayed out when scan sequence is running. If the grayed out key is pressed, an advisory
message is generated. If the equivalent SCPI command is sent, this same message is generated as
part of a “–221, Settings conflict” warning.
Discrete Scan Type does not appear in CXA. The key is gray out and sending the SCPI command will
return Illegal Parameter Error.
Time Domain appears in models that have a DP2 or B40 license and the TDS license installed. The
key is grayed out in all other models, and sending the SCPI command will return an Illegal Parameter
Error.
Preset SMOoth
EMI Receiver Mode Reference 481