10 Monitor Spectrum Measurement
Meas Setup
Key Path
Meas Setup, Signal List, Sort Signals
Remote Command
:CALCulate:SLISt:SORT:AUTO ON|OFF|1|0
:CALCulate:SLISt:SORT:AUTO?
Example CALC:SLIS:SORT:AUTO 1 Turn on auto sorting
CALC:SLIS:SORT:AUTO?
Notes If Scan Sequence is set to Remeasure, signal in the list will only be auto sorted when Remeasure is
completed.
Preset ON
State Saved Saved in instrument state
Range On|Off
Initial S/W Revision A.13.00
Comment
Sets an ASCII comment field for each signal in the list, which will be shown in the “Comment” column of
signal list. This value will be stored in signal list exported file.
Key Path
Meas Setup, Signal List
Remote Command
[:SENSe]:SLISt:COMMent <integer>,“string”
[:SENSe]:SLISt:COMMent? <integer>
Example :SLIS:COMM 2, "unknown"
Preset “” (null String), not affected by Mode Preset, preset by Restore Mode Defaults.
State Saved Saved in instrument state
Initial S/W Revision A.13.00
Limits
Opens the Limit Lines menu that contains keys that control limit lines of the current measurement. Only
the upper limit is used in the Monitor Spectrum measurement.
Key Path
Meas Setup
State Saved Saved in instrument state
Initial S/W Revision A.13.00
Select Limit
Specifies the selected limit. The term “selected limit” is used throughout this document to specify which
limit will be affected by the functions.
EMI Receiver Mode Reference 789