EasyManuals Logo

Keysight x-series User Reference Guide

Keysight x-series
1138 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #117 background imageLoading...
Page #117 background image
3 Programming the Analyzer
STATus Subsystem
Mode All
Remote Command
:STATus:QUEStionable:CALibration:ENABle <integer>
:STATus:QUEStionable:CALibration:ENABle?
Example STAT:QUES:CAL:ENAB 16384 Can be used to query if an alignment is needed, if you have turned off
the automatic alignment process.
Min 0
Max 32767
Status Bits/OPC
dependencies
Sequential command
Initial S/W Revision Prior to A.02.00
Questionable Calibration Event Query
This query returns the decimal value of the sum of the bits in the Questionable Calibration Event register.
The register requires that the associated PTR or NTR filters be set before a condition register bit can set a
bit in the event register. The data in this register is latched until it is queried. Once queried, the register is
cleared.
Mode All
Remote Command
:STATus:QUEStionable:CALibration[:EVENt]?
Example STAT:QUES:CAL?
Preset 0
Status Bits/OPC
dependencies
Sequential command
Initial S/W Revision Prior to A.02.00
Questionable Calibration Negative Transition
This command determines which bits in the Questionable Calibration Condition register will set the
corresponding bit in the Questionable Calibration Event register when the condition register bit has a
negative transition (1 to 0). The variable <integer> is the sum of the decimal values of the bits that you want
to enable.
Mode All
Remote Command
:STATus:QUEStionable:CALibration:NTRansition <integer>
:STATus:QUEStionable:CALibration:NTRansition?
Example STAT:QUES:CAL:NTR 16384 Alignment is not required.
Preset 0
Min 0
EMI Receiver Mode Reference 117

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals