3 Programming the Analyzer
STATus Subsystem
Questionable Power Negative Transition
This command determines which bits in the Questionable Power Condition register will set the
corresponding bit in the Questionable Power Event register when the condition register bit has a negative
transition (1 to 0). The variable <integer> is the sum of the decimal values of the bits that you want to
enable.
Mode All
Remote Command
:STATus:QUEStionable:POWer:NTRansition <integer>
:STATus:QUEStionable:POWer:NTRansition?
Example STAT:QUES:POW:NTR 32 50 MHz Input Power became OK for Cal will be reported to the Power
Summary of the Status Questionable register.
Preset 0
Min 0
Max 32767
Status Bits/OPC
dependencies
Sequential command
Initial S/W Revision Prior to A.02.00
Questionable Power Positive Transition
This command determines which bits in the Questionable Power Condition register will set the
corresponding bit in the Questionable Power Event register when the condition register bit has a positive
transition (0 to 1). The variable <integer> is the sum of the decimal values of the bits that you want to
enable.
Mode All
Remote Command
:STATus:QUEStionable:POWer:PTRansition <integer>
:STATus:QUEStionable:POWer:PTRansition?>
Example STAT:QUES:POW:PTR 32 50 MHz Input Power became too high for Cal will be reported to the
Power Summary of the Status Questionable register.
Preset 32767
Min 0
Max 32767
Status Bits/OPC
dependencies
Sequential command
Initial S/W Revision Prior to A.02.00
Questionable Temperature Register
"Questionable Temperature Condition" on page 137
136 EMI Receiver Mode Reference

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals