EasyManuals Logo

Keysight x-series User Reference Guide

Keysight x-series
1138 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #576 background imageLoading...
Page #576 background image
8 Frequency Scan Measurement
Sweep/ Control
Remote Command
INITiate[1]:CONTinuous OFF
Example INIT:CONT 0; INIT:IMM
Dependencies This key is grayed out when a scan sequence is running. If the grayed out key is pressed, an advisory
message is generated. If the equivalent SCPI command is sent, this same message is generated as
part of a “–221, Settings conflict” warning.
Couplings Meters will do a single measurement and stop. You can start the meters by changing the setting to
continuous, or by sending the remote command INIT:IMM .
Initial S/W Revision A.07.00
Continuous (Meters)
Performs a continuous measurement at meters.
Key Path
Sweep/Control
Remote Command
INITiate[1]:CONTinuous ON
Example INIT:CONT 1
Dependencies This key is grayed out when a scan sequence is running. If the grayed out key is pressed, an advisory
message is generated. If the equivalent SCPI command is sent, this same message is generated as
part of a “–221, Settings conflict” warning.
Couplings Meters will do measurements continuously, until you change the setting to single or a scan sequence
is started.
Initial S/W Revision A.07.00
Restart Meters (Remote Command Only)
Allows you to restart the meters measurement when set to single.
Remote Command
INITiate[1]:IMMediate
INITiate[1]:RESTart
Example
INIT:IMMINIT:REST
Notes This command is remote language compatible with ESU.
Initial S/W Revision A.07.00
576 EMI Receiver Mode Reference

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals