9 Strip Chart Measurement
Meas
Sample Trace Data - Not Constant Envelope
(See below for explanation of variables.)
<soffset> - start offset is an optional real number. (It is in seconds for time-domain traces, and is a
dimensionless index 0 to Npoints 1, for frequency-domain traces). It specifies the amount of data at the
beginning of the trace that will be ignored before the decimation process starts. It is the time or
frequency change from the start of the trace to the point where you want to start using the data. The
default value is zero.
<length> - is an optional real number. (It is in seconds for time-domain traces, and is a dimensionless
index 0 to Npoints 1, for frequency-domain traces). It defines how much data will be compressed into
one value. This parameter has a default value equal to the current trace length.
<roffset> - repeat offset is an optional real number. (It is in seconds for time-domain traces, and is a
dimensionless index 0 to Npoints 1, for frequency-domain traces). It defines the beginning of the next
field of trace elements to be compressed. This is relative to the beginning of the previous field. This
parameter has a default value equal to the <length> variable. Note that this parameter is used for a
completely different purpose when curve fitting (see CFIT above).
<rlimit> - repeat limit is an optional integer. It specifies the number of data items that you want returned.
It will ignore any additional items beyond that number. You can use the Start offset and the Repeat limit
to pick out exactly what part of the data you want to use. The default value is all the data.
EMI Receiver Mode Reference 651

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals