EasyManuals Logo

Keysight x-series User Reference Guide

Keysight x-series
1138 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #681 background imageLoading...
Page #681 background image
9 Strip Chart Measurement
Restart
Even when set for Single operation, multiple sweeps may be taken when Restart is pressed (for example,
when averaging/holding is on). Thus when we say that Restart "restarts a measurement," we may mean:
It restarts the current sweep
It restarts the current measurement
It restarts the current set of sweeps if any trace is in Trace Average, Max Hold or Min Hold
It restarts the current set of measurements if Averaging, or Max Hold, or Min Hold is on for the
measurement
depending on the current settings.
With Average/Hold Number (in Meas Setup menu) set to 1, or Averaging off, or no trace in Trace Average
or Hold, a single sweep is equivalent to a single measurement. A single sweep is taken after the trigger
condition is met; and the analyzer stops sweeping once that sweep has completed. However, with
Average/Hold Number >1 and at least one trace set to Trace Average, Max Hold, or Min Hold (SA
Measurement) or Averaging on (most other measurements), multiple sweeps/data acquisitions are taken
for a single measurement. The trigger condition must be met prior to each sweep. The sweep is stopped
when the average count k equals the number N set for Average/Hold Number. A measurement average
usually applies to all traces, marker results, and numeric results; but sometimes it only applies to the
numeric results.
Once the full set of sweeps has been taken, the analyzer will go to idle state. To take one more sweep
without resetting the average count, increment the average count by 1, by pressing the step up key while
Average/Hold Number is the active function, or sending the remote command CALC:AVER:TCON UP.
EMI Receiver Mode Reference 681

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)