11 APD Measurement
AMPTD Y Scale
Key Path
AMPTD Y Scale
Scope Meas Global
Remote Command
[:SENSe]:POWer[:RF]:GAIN[:STATe] OFF|ON|0|1
[:SENSe]:POWer[:RF]:GAIN[:STATe]?
Dependencies Preamp is not available on all hardware platforms. If the preamp is not present or is unlicensed, the
key is not shown.
The preamp is not available when the electronic/soft attenuator is enabled.
Couplings The act of connecting the U7227A USB Preamplifier to one of the analyzer’s USB ports will cause
the Internal Preamp to be switched on. When this happens an informational message will be
generated: Internal Preamp turned on for optimal operation with USB Preamp.” Note that if the
Internal Preamp was already on, there will be no change to the setting, but if it was Off it will be
switched On, to Full Range.
Note that this same action occurs when the SA mode is selected while the USB Preamp is
connected to one of the analyzer’s USB ports, if it is the first time that the SA mode has run since
powerup, or if the last time the SA mode was running the USB Preamp was NOT connected.
Subsequently disconnecting the USB Preamp from USB does not change the Internal Preamp setting
nor restore the previous setting.
Preset OFF
State Saved Saved in instrument state
Initial S/W Revision Prior to A.02.00
Modified at S/W Revision A.14.00
Key Path
AMPTD Y Scale, Internal Preamp
Scope Meas Global
Remote Command
[:SENSe]:POWer[:RF]:GAIN:BAND LOW|FULL
[:SENSe]:POWer[:RF]:GAIN:BAND?
Dependencies Preamp is not available on all hardware platforms. If the preamp is not present or is unlicensed, the
key is not shown.
If a POW:GAIN:BAND FULL command is sent when a low band preamp is available, the preamp band
parameter is to LOW instead of FULL, and an "Option not installed" message is generated.
Preset LOW
State Saved Saved in instrument state
Initial S/W Revision Prior to A.02.00
Off
Turns the internal preamp off
906 EMI Receiver Mode Reference

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals