[by load]: 0.00001…999999 <kg>, t, lb, oz, g,
mg
— Calibrated at Display only
— Sensitivity (µV/d) Display only
— Test Determine test value
— Exit calibration Save or discard changes due to new calibration.
— Modify May be used only for minor changes (e.g.
changing the dead load, adapting mV/V values
for dead load and/or Max). Otherwise, always
use [New]!
— Param Parameter settings
— Measuretime Input of the measuring period: 5, 10, 20, 40, 80,
160, <320>, 640, 960, 1280, 1600 ms
— Digital filter Selection of the digital filters: <off>, Bessel,
aperiod., Butterw., Tcheby.
— External supply Selection: below or equal 8 V (≤8 V), <above
8 V> (>8 V)
— Fcut Input of cut off frequency, only unless filter not
"off", 0.1–80.0 Hz
— Test mode Selection for display of the deviation from the
test value: <Absolute>, relative
— W&M Selection legal-for-trade mode: <none>, OIML
(impossible when [Range mode] "Multi-
interval" has been selected or Max has more
than 3 decimals), NSC, NTEP
— Standstill time Input of the standstill period:
0.01 s…<0.50 s>…2.0 s (The range depends on
the measuring time.)
— Standstill range Input of the standstill range:
0.00 d…<1.00 d>…10.00 d
— Tare timeout Input of timeout when there is no standstill:
0.1 s…<2.5 >…25 s
— Zeroset range ± Range of zero point when there is no
standstill. Input:
0.00 d…<50.00 d>…10000.00 d
— Zerotrack indic. range Input of the zerotrack indication range:
0.00 d…<0.25 d>…10000.00 d
— Zerotrack step Input of the zerotrack step:
0.00 d…<0.25 d>…10.00 d
— Zerotrack time Input of the zerotrack time: <0.0 s>…25 s
— Overload Input of the weighing range above the
maximum load (Max) without error message:
0…999999 d
— Minimum weight Input of the minimum load:
0 d…<50 d>…999999 d
— Range mode Range selection: <Single range>, Multiple
range, Multi-interval
See also Chapters 7.14.15.2 and 7.14.15.3.
— Range limit 1 Input of the limit 1: In weight, unit same as Max,
transition from small to medium scale interval
Only for [Multiple range] or [Multi-interval]!
7 Getting started X3 Process Indicator PR 5410