[by load]: 0.00001…999999 <kg>, t, lb, oz, g,
mg
Calibrated at Display only
Sensitivity (µV/d) Display only
Test Determine test value
Exit calibration Save or discard changes due to new calibration.
Modify May be used only for minor changes (e.g.
changing the dead load, adapting mV/V values
for dead load and/or Max). Otherwise, always
use [New]!
Param Parameter settings
Measuretime Input of the measuring period: 5, 10, 20, 40, 80,
160, <320>, 640, 960, 1280, 1600 ms
Digital filter Selection of the digital filters: <off>, Bessel,
aperiod., Butterw., Tcheby.
External supply Selection: below or equal 8 V (≤8 V), <above
8 V> (>8 V)
Fcut Input of cut off frequency, only unless filter not
"off", 0.1–80.0 Hz
Test mode Selection for display of the deviation from the
test value: <Absolute>, relative
W&M Selection legal-for-trade mode: <none>, OIML
(impossible when [Range mode] "Multi-
interval" has been selected or Max has more
than 3 decimals), NSC, NTEP
Standstill time Input of the standstill period:
0.01 s…<0.50 s>…2.0 s (The range depends on
the measuring time.)
Standstill range Input of the standstill range:
0.00 d…<1.00 d>…10.00 d
Tare timeout Input of timeout when there is no standstill:
0.1 s…<2.5 >…25 s
Zeroset range ± Range of zero point when there is no
standstill. Input:
0.00 d…<50.00 d>…10000.00 d
Zerotrack indic. range Input of the zerotrack indication range:
0.00 d…<0.25 d>…10000.00 d
Zerotrack step Input of the zerotrack step:
0.00 d…<0.25 d>…10.00 d
Zerotrack time Input of the zerotrack time: <0.0 s>…25 s
Overload Input of the weighing range above the
maximum load (Max) without error message:
0…999999 d
Minimum weight Input of the minimum load:
0 d…<50 d>…999999 d
Range mode Range selection: <Single range>, Multiple
range, Multi-interval
See also Chapters 7.14.15.2 and 7.14.15.3.
Range limit 1 Input of the limit 1: In weight, unit same as Max,
transition from small to medium scale interval
Only for [Multiple range] or [Multi-interval]!
Range limit 2
7 Getting started X3 Process Indicator PR 5410
Minebea Intec EN-155

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Minebea Intec X3 and is the answer not in the manual?

Minebea Intec X3 Specifications

General IconGeneral
BrandMinebea Intec
ModelX3
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals