[Zeroset range]
Define a ±range around the zero point determined by the dead load during calibration;
within this range
- the displayed gross weight can be set to zero by pressing the zero-setting key (or by
a corresponding external command), and
- automatic zero tracking is active.
Setting range: 0.00…10000.00 d
For use in legal metrology a value ≤2% of the Max must be entered, example: 60 d for
3000 e of class III.
[Zerotrack indic. range]
Indication range within which automatic zero tracking compensates deviations.
Setting range: 0.25…10000.00 d
In "legal-for-trade" mode a value of <0.5 d has to be entered.
[Zerotrack step]
If a weight change exceeds the adjusted value, automatic tracking does not function any
more.
Setting range for automatic tracking increments: 0.25…10 d
In "legal-for-trade" mode a value of ≤0.5 d has to be entered.
[Zerotrack time]
Time interval for automatic zero tracking.
Setting range: 0.1…25 s
At 0.0 s the zero tracking is switched off.
For use in legal metrology a value of 1 s must be entered.
[Overload]
Weighing range above the maximum load (Max) without error message.
Setting range: 0…9999999 d
For use in legal metrology a value of max. 9 d = e must be set.
[Minimum weight]
Minimum weight at which a print command can be triggered.
Setting range: 0…9999999 d
For use in legal metrology a value of at least 20 d must be set.
[Range mode]
Selection: <Single range>, Multiple range, Multi-interval
For scale range selection, see Chapter 7.14.15.2 and 7.14.15.3.
Press the
softkey to exit the menu and to save the settings.