— 9-36 —
P IN: IN-position Limit
Sets the criterion for outputting the In-position signal (In-position limit).
◊ Refer to “7.2.6. In-position Limit: Parameter IN for details.
The output IPOS (In-position) activates when the absolute values of the position error counter is
stable within the setting range of the parameter IN (In-position limit) for the time specified in the
parameter IS (In-position stability timer).
Category Parameter
Format
IN data
Data range 0 to 2 621 439 [pulse]
Shipping set 400
IO0: Input/Output Monitor 0 (Electrical condition)
Monitors electrical conditions of the connector CN2 (Input/output signal connector).
◊ Indicates the condition by zeros (0) and ones (1) to each signal port.
◊ Refer to ”7.3.1.1. Electric Condition Monitor: IO0” for details.
Category Monitor
Format
IO0
0
: Input OFF , Output Open
Data range
1
: Input ON, Output Closed
IO1: Input/Output Monitor 1 (Internal recognition)
Monitors the application state of function to the connector CN2 (Control inputs and outputs
signal connector): the polarity reverse and the anti-chattering timer to the control inputs, and the
stability timer for the control outputs.
◊ Indicates the condition by zeros (0) and ones (1) to each signal port.
◊ Refer to ”7.1.3.2. Monitor for Internal Recognition of Input and Output State: IO1” for
details.
Category Monitor
Format
IO1
0
: Input [Not functioning ], Output [Not detect the state]
Data range
1
: Input [Functioning], Output [Detect the state]