4535 612 10261 HD11 Service Manual Page 142
CSIP Level 1 Performance Tests: Test and Inspection Matrices
E:
(E-box, Power Supply)
Yes Yes Whenever any component in the E-box or the power supply has
been serviced.
Whenever any cable attached to the E-box or the power supply
has been serviced.
During preventive maintenance or if configuration is lost, test the
battery.
C:
(Cart)
Yes Yes Whenever any part on the cart has been serviced (a part that is
not a part of one of the other test blocks).
For example: When the power cord, wheels, cart, keyboard,
monitor, rear door, monitor twivel, or the side panels are serviced.
NOTE: Stand-alone, nonimaging and imaging transducers that are
sent to a customer by themselves (not with a system), do not require
installation and therefore do not require this test.
Table 6-3 When to Perform Tests—Details (Continued)
Test Block Name
(Symbol)
After a
PM Task
After an
Installation
or Upgrade Other Situations Requiring Testing