4535 612 10261 HD11 Service Manual Page 141
CSIP Level 1 Performance Tests: Test and Inspection Matrices
Sbf:
(Safety—Transthoracic
BF type transducers)
(See the
Other
Situations
Requiring
Te s t i n g
column)
No This test is only performed if, during the visual inspection of a
transthoracic “BF” type transducer, a physical defect is noted that
could compromise the electrical integrity of the transducer.
For example: A small crack or hole in the lens or body of the
transducer is present.
Tn:
(Non-Imaging
Transducers)
Yes Yes
See NOTE on
page 142.
Whenever just the nonimaging transducer has been serviced.
P:
(Printers)
Yes Yes • Whenever any system or stand-alone printer has been serviced.
• Whenever any circuitry or cables that control or send signals to a
printer have been serviced.
V:
(VCRs)
Yes Yes • Whenever any system or stand-alone VCR has been serviced.
• Whenever any circuitry or cables that control signals to a VCR
have been serviced.
Xp:
(PC)
Yes Yes • Whenever the PC or any of its components has been serviced.
• Whenever any cables connected to the PC have been serviced.
X:
(Software)
No Yes This test is performed during an installation or a repair service only if
system software was loaded. This test ensures that you have loaded
the correct software version.
Table 6-3 When to Perform Tests—Details (Continued)
Test Block Name
(Symbol)
After a
PM Task
After an
Installation
or Upgrade Other Situations Requiring Testing