4535 612 10261 HD11 Service Manual Page 140
CSIP Level 1 Performance Tests: Test and Inspection Matrices
S:
(Safety—System or
Stand-alone Peripherals)
Yes No In any situation where the CE (or customer) suspects that the
electrical safety of the system may have been compromised.
Whenever the ECG/Physio board is changed.
System Safety not required after software-only installation or
upgrades.
Ste:
(Safety—TEE
and Endocavity
Transducers)
Ye s No In any situation where the CE suspects that the electrical safety of
the transducer may have been compromised.
Sep:
(Safety—Epiaortic/
Interoperative
Transducers—CF Type)
Ye s No In any situation where the CE suspects that the electrical safety of
the transducer may have been compromised.
Ti:
(Imaging Transducers)
Ye s Yes
See NOTE on
page 142.
Whenever just the imaging transducer has been serviced.
Table 6-3 When to Perform Tests—Details (Continued)
Test Block Name
(Symbol)
After a
PM Task
After an
Installation
or Upgrade Other Situations Requiring Testing