12 Injection Port
12.1 Packed Column Injection Port (Dual INJ)
123
GC-2014
12.1.2.3 PF menu
PF menu Description
Reference
section
Column
Sets inner diameter, length, and film thickness of a column used.
Values entered here are simply for notes and they are not used for flow
rate calculations.
12.1.3
On/Off
Set the flow controller to be used to “On.”
When the GC starts, the AFC in the current analytical line which is set to be
“On” is controlled.
――
L Flow prog. Sets the flow rate program for the flow controller on the left inlet.
12.1.4
R Flow prog. Sets the flow rate program for the flow controller on the right inlet.
Offset
Performs offset calibration of the AFC sensor. This calibration contributes
to good reproducibility of results.
12.4
Next Flow
Toggles among the AFC set up screens when two or more AFCs are
installed in the system. The [FLOW] key can also be used to switch
screens.
――