12 Injection Port
12.1 Packed Column Injection Port (Dual INJ)
123
GC-2014
12.1.2.3 PF menu
PF menu Description
Reference
section
Column
Sets inner diameter, length, and film thickness of a column used.
Values entered here are simply for notes and they are not used for flow
rate calculations.
12.1.3
On/Off
Set the flow controller to be used to “On.”
When the GC starts, the AFC in the current analytical line which is set to be
“On” is controlled.
――
L Flow prog. Sets the flow rate program for the flow controller on the left inlet.
12.1.4
R Flow prog. Sets the flow rate program for the flow controller on the right inlet.
Offset
Performs offset calibration of the AFC sensor. This calibration contributes
to good reproducibility of results.
12.4
Next Flow
Toggles among the AFC set up screens when two or more AFCs are
installed in the system. The [FLOW] key can also be used to switch
screens.
――

Table of Contents

Other manuals for Shimadzu GC-2014

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu GC-2014 and is the answer not in the manual?

Shimadzu GC-2014 Specifications

General IconGeneral
TypeGas Chromatograph
DisplayLCD
Maximum Heating Rate120°C/min
Data AcquisitionDigital
Maximum Number of Detectors4
WeightApprox. 45 kg
Carrier GasHelium, Nitrogen, Hydrogen
Detector OptionsFID, TCD, ECD, FPD
Injection PortSplit/Splitless, On-column
Gas ControlElectronic Flow Control (EFC)
Temperature Programming Rate0.1 to 120 °C/min
Power Requirements230 V AC, 50/60 Hz
Detector Temperature RangeUp to 400°C (depending on detector type)

Related product manuals