275
17 Printing
17.2
17.
GC-2014
17.2AOC commands
When you press [Print] from the AOC, all items are printed as command names. The list
below shows correspondence of command names to item names.
The setup values, such as “Fast” and “Yes/No” are printed as numbers. Correspondence of
setup values to numbers is shown in the range and default columns in the list below.
Q Command list
Command
name
Item Range Default
WRPT Sample Wash 0−99 2
WMOD Solvent Wash 0−99 1
REPT Number of Injection 1−99 1
IVOL Sample Size 0.1−8.0 1.0
WPRS Pre solvent Wash 0−99 0
PUMP Pumping 0−99 5
WTPP Viscosity 0.0−99.9 0.2
WAIT Dwell Time 0.0−99.9 0
ISPD Inj. Speed (Plunger) Slow: 0, Fast: 2 Fast: 2
SSPD Inj. Speed (Syringe) Slow: 0, Fast: 1 Fast: 1
SAND Inj. Mode 0−40
SOLV Solvent selection
All: 0, only A: 1,
only B: 2, only C: 3
All: 0
SINT Priority Sample No. 1 0
*SSNO
Injected sample No. (Only this sample is
analyzed.)
10
*SNO2 Injected sample No. (for sub AOC) 1 0
AAIR Air Aspiration No: 0, Yes: 1 No: 0
USPD Plunger Aspiration Speed Slow: 0, Middle : 1, Fast: 2 Fast: 2
DSPD Speed of Plunger Slow: 0, Middle : 1, Fast: 2 Fast: 2
HIGH
Syringe Height ( ↑ )
0−20 0
LOWS
Syringe Height ( ↓ )
1.5 ml vial: 0−2
4 ml vial: 0−10
0
INJH Syringe Height (Inj) 0−22 0
STRI Multi - Inj 1−99 1
*FSAM
Final sample No. (Samples after that are not
analyzed.)
10
*WKEY
Washing with solvent before injection in
solvent flush mode
0, 1 0
*UVOL Aspiration volume during pumping 8 µl: 0, 6 µl: 1 8 µl: 0
SLMD Using 3 Solvent Vials No: 0, Yes: 1 No: 0
VIAL Vial size 1.5 ml:0, 4 ml:1 1.5 ml:0
*CKTR With/without tray check With: 0, Without: 1 Without: 0
*TANL Analysis time 0−655 0
*TSTR Analysis start time 0.0−99.9 0.0
LSYR Syringe Volume
10 µl: 0, 50 µl: 1,
250 µl: 2
10 µl: 0