15 Optional Devices
15.5 Setting the CRG Parameters
210
GC-2014
15.5.2 Parameter list
COLUMN OVEN
VALVE Selection: On/Off, Default: Off
CRG Selection: Use/Not Use, Default: Not Use
When the CRG is connected to the column oven, the column oven temperature
control range is extended to -99.0 +400.0 °C from (room temperature + 4 °C)
+400.0 °C.
When using the CRG, set “VALVE” to “On” and “CRG” to “Use”.
When using valve control only, set “VALVE” to “On” and set “CRG” to “Not Use”.
INJ 2
VALVE Selection: On/Off, Default: Off
CRG Selection: Use/ Not Use, Default: Not Use
When the PTV is installed on INJ2, the injection port temperature control range is
extended to -99.0+400.0 °C from 0.0400.0 °C
When using the CRG, set “VALVE” to “On” and “CRG” to “Use”.
When using valve control only, set “VALVE” to “On” and set “CRG” to “Not Use”.
15.5.3 PF menu
PF menu Description
Reference
section
Next
Toggles the screen among AOC parameters AUX temperature AUX
APC AUX AMC CRG screens in this order.
――

Table of Contents

Other manuals for Shimadzu GC-2014

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu GC-2014 and is the answer not in the manual?

Shimadzu GC-2014 Specifications

General IconGeneral
TypeGas Chromatograph
DisplayLCD
Maximum Heating Rate120°C/min
Data AcquisitionDigital
Maximum Number of Detectors4
WeightApprox. 45 kg
Carrier GasHelium, Nitrogen, Hydrogen
Detector OptionsFID, TCD, ECD, FPD
Injection PortSplit/Splitless, On-column
Gas ControlElectronic Flow Control (EFC)
Temperature Programming Rate0.1 to 120 °C/min
Power Requirements230 V AC, 50/60 Hz
Detector Temperature RangeUp to 400°C (depending on detector type)

Related product manuals