16 Special Functions
16.1 Time Scheduler
215
GC-2014
16.1.8 Parameters available in the time scheduler
16.1.9 Changing schedule parameters
To change the contents of an existing time schedule, move the cursor using the [ ] and [ ]
keys to select the schedule to be changed, and press the [ENTER] key.
To change the time:
When the Time Schedule select screen appears, enter new numeric values and press
[Finish] (PF menu), if you would like to change only the time.
To change the parameters:
Move the cursor, change the parameter, and then press [Finish] (PF menu) to complete
the change.
To cancel any changes:
If you have changed the schedule but would like to return to the former schedule, press
[Cancel] (PF menu) before pressing [Finish] (PF menu). The schedule returns to its former
status.
Parameter Description
Stop
NUMBER
OF
TIMES
OF RUN
Time
scheduler
stop
Specifies the number of times the program can execute before
the schedule stops or switches to another file. If greater than 2,
The choices are “Continuous” or “Pause at each run”. For an
AOC, select “Continuous”. When “Pause at each run” is
chosen, the program executes at set intervals continuously at
set intervals.
If another program (such as a temperature program) is running
when time schedule execution time expires, priority is given to
the running program. When the current program finishes, the
time schedule stops or file is changed to another file.
Schedule
0-9 Load
GC Start/Stop Sets start/stop of gas chromatograph.

Table of Contents

Other manuals for Shimadzu GC-2014

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu GC-2014 and is the answer not in the manual?

Shimadzu GC-2014 Specifications

General IconGeneral
TypeGas Chromatograph
DisplayLCD
Maximum Heating Rate120°C/min
Data AcquisitionDigital
Maximum Number of Detectors4
WeightApprox. 45 kg
Carrier GasHelium, Nitrogen, Hydrogen
Detector OptionsFID, TCD, ECD, FPD
Injection PortSplit/Splitless, On-column
Gas ControlElectronic Flow Control (EFC)
Temperature Programming Rate0.1 to 120 °C/min
Power Requirements230 V AC, 50/60 Hz
Detector Temperature RangeUp to 400°C (depending on detector type)

Related product manuals