13 Detector
13.1 Hydrogen Flame Ionization Detector (FID)
161
GC-2014
13.1.5 Igniting and Extinguishing the FID
13.1.5.1 Confirmation prior to FID ignition
Confirm the following points prior to FID ignition.
1. Connect the column.
2. Specify the FID to be ignited in a flow line. (Refer to “8.3 Specifying the Analytical
Flow Line Components ([Line Config])”)
3. Set the FID to be ignited to On. (Refer to “13.1.2 Setting the Detector.”)
4. Turn the main valve of the gas cylinder to supply gas.
5. Set the flow rate of the detector gas. For APC, press [On/Off] (PF menu) on the
screen to set the detector gas so that all gas control is turned on.
(Refer to “13.1.3 Setting the Detector Gas (manual flow controller)” or “13.1.4
Setting the detector gas flows(APC)”.)
13.1.5.2 FID ignition procedures
Q Automatic ignition
The system starts and the FID is ignited automatically.
1. Press the [SYSTEM] key and set the following parameters. Change other
parameters if necessary.
2. Start the GC.
GC starts when [Start GC] (PF menu) is pressed in manual starting and when the
power is turned on in automatic starting.
START TEMP/DET   Yes
DETECTOR   On
AUTO IGNITE   On
(a) Main screen (b) [Start Seq] (PF menu)
Fig. 13.1.6   Setting starting procedures (automatic ignition)