13 Detector
13.1 Hydrogen Flame Ionization Detector (FID)
162
GC-2014
3. After the GC starts, gas control starts. Temperature/detector control starts after the
start time passes. When the detector’s temperature becomes READY, the FID is
ignited automatically.
Q Manual ignition
The FID is ignited by key operation.
1. Press the [SYSTEM] key and set the following parameters. Change other
parameters if necessary.
2. Start the GC.
GC starts when [Start GC] (PF menu) is pressed in manual starting and when the
power is turned on in automatic starting.
3. After the GC starts, gas control starts. Temperature/detector control starts after the
start time passes.
4. After the detector’s temperature becomes 100 °C or higher, press [Ignite
] (PF
menu) on the [DET] key main screen.
START TEMP/DET   Yes
DETECTOR   On
AUTO IGNITE   Off
(a) Main screen (b) [Start Seq] (PF menu)
Fig. 13.1.7   Setting starting procedures (manual ignition)