18 Error Messages
18.1 Error Messages
297
GC-2014
18.1.7 Operations errors
Value outside of range was entered
When the numeric value entered is out of the valid range, this message or valid setting
range is displayed. Enter a valid number.
File operation errors
These massages appear when an incorrect file operation was attempted.
Press another key, and continue operation.
Overflow of calculated pressure value
The carrier gas pressure is calculated from the linear velocity, flow rate, or the split ratio. The
value you have input is outside the set range. Change the conditions so that the pressure is
within the set range, and enter a new value.
Overflow of calculated flow rate value
The carrier gas total flow rate is calculated from the split ratio or the pressure. The value you
have input is outside the set range. Change the conditions and enter a new value.
Code Message CS
5001 Input parameter out of range No
Code Message CS
5002 Invalid file no. No
5003 This file is now used No
Code Message CS
5006 CAR1 calc. prss out of range No
5007 CAR2 calc. prss out of range No
5008 CAR3 calc. prss out of range No
Code Message CS
5009 CAR1 calc. flow out of range No
5010 CAR2 calc. flow out of range No
5011 CAR3 calc. flow out of range No

Table of Contents

Other manuals for Shimadzu GC-2014

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu GC-2014 and is the answer not in the manual?

Shimadzu GC-2014 Specifications

General IconGeneral
TypeGas Chromatograph
DisplayLCD
Maximum Heating Rate120°C/min
Data AcquisitionDigital
Maximum Number of Detectors4
WeightApprox. 45 kg
Carrier GasHelium, Nitrogen, Hydrogen
Detector OptionsFID, TCD, ECD, FPD
Injection PortSplit/Splitless, On-column
Gas ControlElectronic Flow Control (EFC)
Temperature Programming Rate0.1 to 120 °C/min
Power Requirements230 V AC, 50/60 Hz
Detector Temperature RangeUp to 400°C (depending on detector type)

Related product manuals