EasyManuals Logo

Shimadzu GC-2014 Instruction Manual

Shimadzu GC-2014
318 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #158 background imageLoading...
Page #158 background image
12 Injection Port
12.3 Direct Injection System
146
GC-2014
12.3.1 Setting the temperature
12.3.1.1 Screen description
Press the [INJ] key to display the injection port main screen shown in Fig. 12.3.2 and to
set the injection port temperature.
12.3.1.2 Parameter list
TEMP
Range: 0.0400.0 °C, Default: 25.0 °C
Set the injection port temperature.
Keep the injection port at the lowest practical temperature for the analysis; this prolongs
the service life of the temperature sensor. To instantly vaporize the injected sample, set
the injection port temperature approximately 30 °C higher than the final column oven
temperature. Because the sample quantity is usually minute, the sample is vaporized at
temperatures lower than the boiling point of sample compounds. To set the maximum
temperature limit, refer to “16.6.4 Setting the maximum temperature limits” in “16.6 GC
Configuration”.
12.3.1.3 PF menu
Fig. 12.3.2   [INJ] key main screen
PF menu Description
Reference
section
Print Prints the temperature program to a Chromatopac.
――
Next Inj
Toggles to the next setup screen when two or more injection ports are
installed in this system.
The [INJ] key can also be used to switch screens.
――

Table of Contents

Other manuals for Shimadzu GC-2014

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu GC-2014 and is the answer not in the manual?

Shimadzu GC-2014 Specifications

General IconGeneral
TypeGas Chromatograph
DisplayLCD
Maximum Heating Rate120°C/min
Data AcquisitionDigital
Maximum Number of Detectors4
WeightApprox. 45 kg
Carrier GasHelium, Nitrogen, Hydrogen
Detector OptionsFID, TCD, ECD, FPD
Injection PortSplit/Splitless, On-column
Gas ControlElectronic Flow Control (EFC)
Temperature Programming Rate0.1 to 120 °C/min
Power Requirements230 V AC, 50/60 Hz
Detector Temperature RangeUp to 400°C (depending on detector type)

Related product manuals