16 Special Functions
16.3 Time Program
229
GC-2014
16.3.7 Time program setup example
Example 1: When both positive and negative peaks appear on the chromatogram
Four minutes after the program execution starts, the polarity of the detector DET
1 (TCD) is switched from plus to minus. Two minutes later (six minutes after the
program execution starts), the polarity is switched back to plus.
First program Screen display
  1. Time: 4.0 min
  2. Detector
  3. Polarity
  4. DET 1
  5. - (minus)
Second program
  6. Time: 6.0 min
     Steps 24 are repeated.
  7. + (plus)
  8. The time program is set up.

Table of Contents

Other manuals for Shimadzu GC-2014

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu GC-2014 and is the answer not in the manual?

Shimadzu GC-2014 Specifications

General IconGeneral
TypeGas Chromatograph
DisplayLCD
Maximum Heating Rate120°C/min
Data AcquisitionDigital
Maximum Number of Detectors4
WeightApprox. 45 kg
Carrier GasHelium, Nitrogen, Hydrogen
Detector OptionsFID, TCD, ECD, FPD
Injection PortSplit/Splitless, On-column
Gas ControlElectronic Flow Control (EFC)
Temperature Programming Rate0.1 to 120 °C/min
Power Requirements230 V AC, 50/60 Hz
Detector Temperature RangeUp to 400°C (depending on detector type)