23
2 Before Use  
2.3
2.
GC-2014
2.3Outputting Digital Signals to a Personal
Computer
GC-2014 can be directly connected to a personal computer to output digital signals.
GCsolution software allows a PC to control the unit and take data.
For operation of GCsolution, refer to its instruction manual.
Connecting the RS-232C cable
Connect the RS-232C cable attached to GCsolution workstation to the connector on the
back of the GC. (Fig. 2.3.1)
To connect one PC with more than one GCs, separate RS-232C cables are necessary and
an expanded COM port needs to be attached to the PC.
AOC built-in power source has the same connector. Connect the cables correctly.
NOTE
ANALOG
OUT 2
ANALOG
OUT 1
AOC
INJECIOR1 INJECIOR2 SAMPLER
EVENT
OPTION
1
2
3
4
5
6
RS
-
232C
RS
-
232C OP1 LINKSTART
OUT/READY IN
START IN
READY OUT
AOC POWER SUPPLY
START OUT
RS-232C
Back of the GC
AOC built-in power source
Fig. 2.3.1   Connection of RS-232C cable

Table of Contents

Other manuals for Shimadzu GC-2014

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu GC-2014 and is the answer not in the manual?

Shimadzu GC-2014 Specifications

General IconGeneral
TypeGas Chromatograph
DisplayLCD
Maximum Heating Rate120°C/min
Data AcquisitionDigital
Maximum Number of Detectors4
WeightApprox. 45 kg
Carrier GasHelium, Nitrogen, Hydrogen
Detector OptionsFID, TCD, ECD, FPD
Injection PortSplit/Splitless, On-column
Gas ControlElectronic Flow Control (EFC)
Temperature Programming Rate0.1 to 120 °C/min
Power Requirements230 V AC, 50/60 Hz
Detector Temperature RangeUp to 400°C (depending on detector type)