16 Special Functions
16.2 Batch Schedule
222
GC-2014
16.2.5 Parameter list
START VIAL NO.
Range:
1
Default: 0
FINAL VIAL NO.
Range:
1
Default: 0
1: With no auto sampler
Short rack ... 16
Long rack ... 112
With auto sampler 1150 (maximum)
When the auto sampler is used, the possible depends on the vial rack type and the number
of racks.
The default, “0”, indicates that the setting is not given.
SAMPLE
Inject sample
The specified sample vial No. is analysed.
Without sample
Vials are not used for samples. Instead samples are introduced from a gas sampler with a
valve.
GC FILE NO.
Range: FILE 09, Default: FILE0
Set the file No. of the GC method to be used.
CHANNEL NO.
Range: Ch1/Ch2, Default: Ch 1
Set the channel No. connecting the GC to the Chromatopac. This setting is ignored when
using GC solution software.
DATA PROCESSOR FILE NO.
Range: FILE 09, Default: FILE0
Specify a date processor (Chromatopac) date integration file. This setting is ignored when
using GC solution software.
NUMBER OF INJECTION
Range: 099, Default: 0
Set the number of times to inject each sample.
SAMPLE TYPE
Unknown Select this item to analyze a sample of unknown concentration.
Analyze a standard and create a calibration curve before analyzing an unknown
sample.
Standard Select this item to analyze a standard of known concentration.
SLOPE TEST
Range: On/Off, Default: Off
When “On” is selected, the slope test is performed before running the batch schedule.
CLEAN UP
Range: On/Off, Default: Off
When “On” is selected, clean up is performed before running the batch schedule.