16 Special Functions
16.6 GC Configuration
251
GC-2014
16.6.10 Temperature offset
If the actual temperature of the various heated zones (as measured by a thermocouple)
differs flow the set temperature, set a temperature offset value to compensate.
16.6.10.1 Screen description
Select “6. GC CONFIGURATION” from the [FUNC] key and select “8. TEMPERATURE
OFFSET” to display the temperature offset screen shown in Fig. 16.6.22.
16.6.10.2 Parameter list
COLUMN/INJ1/INJ2/DET1/DET2/AUX3/AUX4/AUX5
Range: -10.00+10.00, Default: 0.00
The temperature of each heated zone can be offset, Uninstalled ports are not displayed.
For example, when “-1 °C” is obtained by subtracting the column oven temperature
measured by thermocouple from the gas chromatograph value with an offset value of
“0”, input “-1” as the offset value.
Fig. 16.6.22   Offset set up screen

Table of Contents

Other manuals for Shimadzu GC-2014

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu GC-2014 and is the answer not in the manual?

Shimadzu GC-2014 Specifications

General IconGeneral
TypeGas Chromatograph
DisplayLCD
Maximum Heating Rate120°C/min
Data AcquisitionDigital
Maximum Number of Detectors4
WeightApprox. 45 kg
Carrier GasHelium, Nitrogen, Hydrogen
Detector OptionsFID, TCD, ECD, FPD
Injection PortSplit/Splitless, On-column
Gas ControlElectronic Flow Control (EFC)
Temperature Programming Rate0.1 to 120 °C/min
Power Requirements230 V AC, 50/60 Hz
Detector Temperature RangeUp to 400°C (depending on detector type)

Related product manuals