145
12 Injection Port  
12.3
12.
GC-2014
12.3Direct Injection System
Q Direct injection system
In a direct injection system, a wide-bore (0.53 mm or greater) column is used. Nearly the
entire amount of sample injected is inrtoduced on the column. This usually results in better
sensitivity than narrow-bore columns with split injections. However, peak shapes are broad,
which decreases resolution and can lead to a higher S/N ratio.
The direct injection system uses a WBI (wide-bore injection) injection port. When WBI is
specified, DIRECT injection mode is automatically used and split mode is not available on
the [INJ] key main screen.
Fig. 12.3.1  
Control
Control
TFC
CARRIER
PURGE
PURGE VENT
Capillary column

Table of Contents

Other manuals for Shimadzu GC-2014

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu GC-2014 and is the answer not in the manual?

Shimadzu GC-2014 Specifications

General IconGeneral
TypeGas Chromatograph
DisplayLCD
Maximum Heating Rate120°C/min
Data AcquisitionDigital
Maximum Number of Detectors4
WeightApprox. 45 kg
Carrier GasHelium, Nitrogen, Hydrogen
Detector OptionsFID, TCD, ECD, FPD
Injection PortSplit/Splitless, On-column
Gas ControlElectronic Flow Control (EFC)
Temperature Programming Rate0.1 to 120 °C/min
Power Requirements230 V AC, 50/60 Hz
Detector Temperature RangeUp to 400°C (depending on detector type)

Related product manuals