EasyManua.ls Logo

Shimadzu GC-2014 - Parameter List; Example: Starting the System with Carrier Gas Flow

Shimadzu GC-2014
318 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7 Starting and Stopping the GC [SYSTEM]
7.3 Specifying Start Procedures
87
GC-2014
7.3.2 Parameter list
FILE LOAD
Selection: File No. 09, Default: Current file
Selects a file to be loaded the next time the power is turned on or the GC restarted.
START GC
Selection: Auto Start/ Manual Start/Semi-Auto, Default: Manual Start
Sets the start method for the next time the power is turned on:
Select “Auto Start” to automatically start the GC.
Select “Manual Start” to start the GC by presssing [Start GC] (PF menu) from the [SYSTEM]
key main screen.
Select “Semi-Auto” to start carrier gas flow only. The GC must still be started by pressing
[Start GC] (PF menu) from the the [SYSTEM] key main screen.
START TIME
Range: 0.06000.0 min, Default: 0.0 min
Sets the period of time after gas control starts until temperature/detector control starts.
The “Start Time” value on the [SYSTEM] key main screen is set here.
The start time set here can be overridden by entering another start time in the [Start GC] (PF menu)
screen.
This function is useful if the preset start time is too long.
START TEMP/DET
Selection: Yes/No, Default: Yes
Select “Yes” to start temperature/detector control after the start time is elapses.
Select “No” to continue carrier gas flow only without starting temperature/detector control.
DETECTOR
Selection: On/Off, Default: On
Prepares the configured detector for analysis, but does not ignite the FID or FPD.
The “DETECTOR” setting on the [SYSTEM] key main screen is set here.
AUTO IGNITE
Selection: On/Off, Default: On
Establishes FID/FPD ignition conditions.
Select “On” for the FID/FPD detector to stand by in the ignited status. Select “Off” for it stand
by without igniting.
The “AUTO IGNIT” setting on the [SYSTEM] key main screen is set here.
CLEAN UP
Selection: Off/Analysis Para/Clean Up Para, Default: Off
When performing clean up, select whether to use an analysis method or the program set in
[Clean Up] (PF menu).
The “CLEAN UP” setting on the [SYSTEM] key main screen is set here.
7.3.3 Example: starting the system with carrier gas flow
In this example, carrier gas flows for a certain period of time before temperature control
begins. The length of time the carrier gas should flow depends on how long the carrier gas
flow was shut off.
If the same column is used in the same analytical flow line as the day before:
set a START TIME of approx. 10 minutes.
If a different column is installed on the same flow line that has been in use:
set a START TIME of approx.10 minutes.
If the system has been out of use for a time with no column connected:
set a START TIME of 1 to several hours.
NOTE

Table of Contents

Other manuals for Shimadzu GC-2014

Related product manuals