15 Optional Devices
15.4 Setting the AUX AMC Parameters
208
GC-2014
15.4.2 Parameter list
FLOW RATE
Range: 0.0100.0 ml/min, Default: 50 ml/min
Set the flow rate when the control mode is set to “FLOW”.
GAS TYPE
Selection: He/N
2/H2/Ar, Default: He
Set the type of gas supplied to the AMC. This parameter is used to calculate the flow rate.
15.4.3 PF menu
PF menu Description
Reference
section
Offset
Performs offset calibration of AMC sensor to improve the reproducibility of
results.
――
Next
Toggles the screen among AOC parameters AUX temperature AUX
APC AUX AMC CRG screens in this order.
3.6
On/Off Set the AMC to be used to “On.” Default is “On.”
――