15 Optional Devices
15.4 Setting the AUX AMC Parameters
208
GC-2014
15.4.2 Parameter list
FLOW RATE
Range: 0.0100.0 ml/min, Default: 50 ml/min
Set the flow rate when the control mode is set to “FLOW”.
GAS TYPE
Selection: He/N
2/H2/Ar, Default: He
Set the type of gas supplied to the AMC. This parameter is used to calculate the flow rate.
15.4.3 PF menu
PF menu Description
Reference
section
Offset
Performs offset calibration of AMC sensor to improve the reproducibility of
results.
――
Next
Toggles the screen among AOC parameters AUX temperature AUX
APC AUX AMC CRG screens in this order.
3.6
On/Off Set the AMC to be used to “On.” Default is “On.”
――

Table of Contents

Other manuals for Shimadzu GC-2014

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu GC-2014 and is the answer not in the manual?

Shimadzu GC-2014 Specifications

General IconGeneral
TypeGas Chromatograph
DisplayLCD
Maximum Heating Rate120°C/min
Data AcquisitionDigital
Maximum Number of Detectors4
WeightApprox. 45 kg
Carrier GasHelium, Nitrogen, Hydrogen
Detector OptionsFID, TCD, ECD, FPD
Injection PortSplit/Splitless, On-column
Gas ControlElectronic Flow Control (EFC)
Temperature Programming Rate0.1 to 120 °C/min
Power Requirements230 V AC, 50/60 Hz
Detector Temperature RangeUp to 400°C (depending on detector type)