203
15 Optional Devices  
15.2
15.
GC-2014
15.2Setting AUX Temperature
Temperature controller used for parts other than the column oven, injection port, and detector
can be set.
15.2.1 Screen description
When an optional device such as heater is installed, press the [OPTION] key to display the
screen shown in Fig. 15.2.1. If a different screen appears, press the [OPTION] key again or
[NEXT] (PF menu) to display the correct screen.
The screen toggles among AOC parameters AUX temperature AUX APC AUX AMC
CRG screens in this order.
15.2.2 Parameter list
TEMPERATURE SETTING VALUE
Range: 0.0400.0 °C, Default: 25.0 °C
15.2.3 PF menu
Fig. 15.2.1   Main screen to set AUX temperature
PF menu Description
Reference
section
Print Prints each temperature through a Chromatopac. 17.1
NEXT
Toggles the screen among AOC parameters AUX temperature AUX
APC AUX AMC CRG screens in this order.
――
On/Off Displays a sub screen to set a used unit to On. 15.2.4