EasyManua.ls Logo

Shimadzu GC-2014 - Communication Errors

Shimadzu GC-2014
318 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
18 Error Messages
18.1 Error Messages
293
GC-2014
Abnormal component operation
Verify whether gas is supplied stably at the required pressure. If there is no problem in the
gas supply, the control system, such as the APC, may have failed. Turn off the system and
contact your Shimadzu representative.
18.1.4 Communication errors
External device communication errors (i.e., Chromatopac)
One of these messages appears during a link failure or communication failure. (When a
communication error occurs, the link is automatically disconnected.)
When one of these messages appears, check the connection status, and reset the link.
Code Message CS
2144 CAR1. L DAFC is out of control Yes
2145 CAR1. R DAFC is out of control Yes
2146 CAR2. L DAFC is out of control Yes
2147 CAR2. R DAFC is out of control Yes
2148 AMC.L is out of control Yes
2149 AMC.L is out of control Yes
2150 AMC.R is out of control Yes
2151 AMC1 is out of control Yes
2152 AMC2 is out of control Yes
2153 AMC3 is out of control Yes
2154 AMC4 is out of control Yes
2155 AMC5 is out of control Yes
2156 AMC6 is out of control Yes
2157 AMC7 is out of control Yes
2158 AMC8 is out of control Yes
2159 AMC9 is out of control Yes
2160 AMC10 is out of control Yes
2161 CAR1 SAFC is out of control Yes
2162 CAR2 SAFC is out of control Yes
Code Message CS
4001 Time out No
4002 Parity error No
4003 Message is not accepted No
4004 Data is invalid No
4005 Command is invalid No
4006 Data is out of range No
4007 TRS port is shut down No
4008 TRS file error No