18 Error Messages
18.1Error Messages
This system is programmed to self-diagnose certain errors. When an error is detected, an
alarm sounds and one of the error messages shown below is displayed. The detected errors
are recorded in the “Error Log”.
For errors of which the “CS” column in the table below shows Yes, a screen to confirm about
resetting them appears. When the confirmation screen appears, select one of the following
actions:
Codes shown in the tables below identify errors. Provide the code when calling service
representative about an error. These codes are also recorded in the error log.
18.1.1 System errors
Power supply failure
If the above error messages appear, the hardware has failed. The GC system cannot be
used in this condition. Turn off the system and contact your Shimadzu representative.
Room temperature sensor/atmospheric pressure sensor error
The room temperature or atmospheric pressure is out of the performance range. If this error
occurs even though the actual room temperature/atmospheric pressure is within the range,
the sensors may have failed. The system cannot be used in this condition. Turn off the
system and contact your Shimadzu representative.
Reset Error Resets the system parameters, restoring the conditions prior to the error.
System control resumes. If the cause of the error has been corrected, the
error does not occur any more.
Ignore Error Select this to keep the system as it is for a while for some reason, for
example, in order to remove the cause of the error.
Code Message CS
0001 DC 5 V is out of range No
0002 DC 24 V is out of range No
0003 DC -15 V is out of range No
Code Message CS
0005 Room temperature is out of range No
0006 Atm. pressure is out of range No

Table of Contents

Other manuals for Shimadzu GC-2014

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu GC-2014 and is the answer not in the manual?

Shimadzu GC-2014 Specifications

General IconGeneral
TypeGas Chromatograph
DisplayLCD
Maximum Heating Rate120°C/min
Data AcquisitionDigital
Maximum Number of Detectors4
WeightApprox. 45 kg
Carrier GasHelium, Nitrogen, Hydrogen
Detector OptionsFID, TCD, ECD, FPD
Injection PortSplit/Splitless, On-column
Gas ControlElectronic Flow Control (EFC)
Temperature Programming Rate0.1 to 120 °C/min
Power Requirements230 V AC, 50/60 Hz
Detector Temperature RangeUp to 400°C (depending on detector type)

Related product manuals