18 Error Messages
18.1 Error Messages
294
GC-2014
AOC communication errors
One of these messages appears during a link failure or communication failure. (When a
communication error occurs, the link is automatically disconnected.)
When one of these messages appears, check the connection status, and reset the link.
18.1.5 Detector errors
Over current
Current of the detector is abnormal. If the resistance of the filament becomes abnormally
high when the TCD or the FTD is in use, the protection circuit is actuated to prevent damage
to the filament, and an alarm sounds.
If either error has occurred, turn off the system.
Possible reasons for the protection circuit to be activated are described below.
• The set current value exceeds the maximum operating current.
• Gas is not flowing.
• Gas is leaking.
• A significant amount of air is present in the flow line (for TCD).
Correct the error then turn on the system. If the system does not recover after several
resets, or the reason for the error cannot be located, contact your Shimadzu representative.
Code Message CS
4009 AOC command is invalid No
4010 AOC data is out of range No
4011 AOC time out No
4012 AOC2 TRS error No
4013 AOC link error No
Code Message CS
4101 DET#1 TCD cell error Yes
4102 DET#2 TCD cell error Yes
4103 DET#3 TCD cell error Yes
4104 DET#4 TCD cell error Yes
4105 DET#1 FTD current error Yes
4106 DET#2 FTD current error Yes
4107 DET#3 FTD current error Yes
4108 DET#4 FTD current error Yes