7 Starting and Stopping the GC [SYSTEM]
7.2 Specifying Clean Up Parameters
85
GC-2014
CLEAN UP TIME
Range: 0.009999.99 min, Default: 0.00 min
Sets the final flow rate hold time for the clean up program.
7.2.3 PF menu
PF menu Description
Reference
section
Temp Prog Sets the column oven temperature for the clean up program. 11.2
Press Prog
Sets the column inlet pressure for the clean up program.
It is displayed when the control mode is “pressure” on the [FLOW] key
screen of the AFC.
12.5.5
Flow Prog
Sets the flow rate for the clean up program.
Sets the column flow rate program for the DAFC and the total flow rate
program for the AFC.
Display is possible for the DAFC and when the control mode is set at “flow
rate” on the [FLOW] key screen of the AFC.
12.5.6
Run
Displayed only while GC is in system ON status. Immediately runs the clean
up program.
――
Stop
Stops clean up.
This item is displayed only when the clean up program is in progress.
――
Next Line Displays the clean up program set up screen for another analytical flow line.
――

Table of Contents

Other manuals for Shimadzu GC-2014

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu GC-2014 and is the answer not in the manual?

Shimadzu GC-2014 Specifications

General IconGeneral
TypeGas Chromatograph
DisplayLCD
Maximum Heating Rate120°C/min
Data AcquisitionDigital
Maximum Number of Detectors4
WeightApprox. 45 kg
Carrier GasHelium, Nitrogen, Hydrogen
Detector OptionsFID, TCD, ECD, FPD
Injection PortSplit/Splitless, On-column
Gas ControlElectronic Flow Control (EFC)
Temperature Programming Rate0.1 to 120 °C/min
Power Requirements230 V AC, 50/60 Hz
Detector Temperature RangeUp to 400°C (depending on detector type)

Related product manuals