18 Error Messages
18.1 Error Messages
301
GC-2014
18.1.9 Warning messages
These messages appear when the time or the count exceeds the set value. Replace the
subject component and reset the count on the menu of the DIAG key screen.
Code 9002 warning requires component replacement and counter reset by a serviceperson.
Contact your Shimadzu representative.
This message appears when the system was started before it was ready. Normally, do not
start analysis until the system is ready.
If this message appears when all the parameters including temperature and flow rate are
ready, check the ready setting for unused components and check the equilibration time.
This message appears when ignition sequence was re-attempted because the ignition
failed. This does not stop the execution of analysis. If this message appears frequently,
check “DETECTOR IGNITE” and “DETECTOR IGNITION” by the standard diagnosis
procedure. Also check for gas leaks and verify the gas flow rates. If set values are correct,
contact your Shimadzu representative.
These messages appear when the sensor use time exceeds the preset value. Component
replacement and reset by a serviceperson are required. Contact your Shimadzu representative.
Code Message CS
9000 COL CRG use time is over Yes
9001 INJ2 CRG use time is over Yes
9002 Fan use time is over Yes
9004 CAR1 septum counter is over No
9005 CAR2 septum counter is over No
9006 CAR3 septum counter is over No
9007 CAR1 insert counter is over No
9008 CAR2 insert counter is over No
9009 CAR3 insert counter is over No
Code Message CS
9010 System is not ready No
Code Message CS
9011 Ignition finished (retried) No
Code Message CS
9012 COL sensor use time is over Yes
9013 INJ1 sensor use time is over Yes
9014 DET1 sensor use time is over Yes
9015 INJ2 sensor use time is over Yes
9016 DET2 sensor use time is over Yes
9017 AUX3 sensor use time is over Yes
9018 AUX4 sensor use time is over Yes
9019 AUX5 sensor use time is over Yes

Table of Contents

Other manuals for Shimadzu GC-2014

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu GC-2014 and is the answer not in the manual?

Shimadzu GC-2014 Specifications

General IconGeneral
TypeGas Chromatograph
DisplayLCD
Maximum Heating Rate120°C/min
Data AcquisitionDigital
Maximum Number of Detectors4
WeightApprox. 45 kg
Carrier GasHelium, Nitrogen, Hydrogen
Detector OptionsFID, TCD, ECD, FPD
Injection PortSplit/Splitless, On-column
Gas ControlElectronic Flow Control (EFC)
Temperature Programming Rate0.1 to 120 °C/min
Power Requirements230 V AC, 50/60 Hz
Detector Temperature RangeUp to 400°C (depending on detector type)

Related product manuals