12 Injection Port
12.2 Split/Splitless Injection System
128
GC-2014
12.2.1.2 Parameter list
TEMP
Range: 0.0400.0 °C, Default: 25.0 °C
Set the temperature of the injection port.
Keep the injection port temperature as low as is practical to increase the service life of
the temperature sensor. In order to instantly vaporize samples inside the injection port,
set the injection port temperature approximately 30 °C above the column oven
temperature (final temp). Due to the minute quantity of sample injected, the sample
vaporizes at temperatures lower than the boiling point of sample elements. To set the
injection port maximum temperature limit, refer to “16.6.4 Setting the maximum
temperature limits” in “16.6 GC Configuration”.
12.2.1.3 PF menu
PF menu Description
Reference
section
Print Prints the injection port temperature on Chromatopac.
――
Next Inj
Toggles among the setup screeens if two or more injection ports are
installed in this system.
The [INJ] key can also be used to switch screens.
――