12 Injection Port
12.5 Setting the Flow Rate Parameters
151
GC-2014
Example 1: Column oven temperature = 50 °C(carrier gas = helium)
Example 2: Column oven temperature = 200 °C (carrier gas = helium)
Fig. 12.5.2 shows the relationship between the column flow rate and the column inlet pressure.
   60 πd
4
   P P0
2
P0
2
Fc      ×         ×10
3
   256 µL        P0
Fc : Column flow rate [ml/min]
d : Column I.D [mm]
L : Column length [m]
P : Column inlet pressure [kPa]
P
0 : Column outlet pressure = atmospheric pressure [kPa]
µ : Viscosity coefficient (19.4 µPa•s (helium, 20 °C))
Fig. 12.5.2  
Length
30 m 60 m
I.D. of column Film thickness
0.25 mm 0.25 µm 100 kPa 210 kPa
0.32 mm 0.25 µm 60 kPa 120 kPa
0.53 mm 1.5 µm 20 kPa 40 kPa
Length
30 m 60 m
I.D. of column Film thickness
0.25 mm 0.25 µm 130 kPa 275 kPa
0.32 mm 0.25 µm 80 kPa 160 kPa
0.53 mm 1.5 µm 30 kPa 60 kPa
NOTE

Table of Contents

Other manuals for Shimadzu GC-2014

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu GC-2014 and is the answer not in the manual?

Shimadzu GC-2014 Specifications

General IconGeneral
TypeGas Chromatograph
DisplayLCD
Maximum Heating Rate120°C/min
Data AcquisitionDigital
Maximum Number of Detectors4
WeightApprox. 45 kg
Carrier GasHelium, Nitrogen, Hydrogen
Detector OptionsFID, TCD, ECD, FPD
Injection PortSplit/Splitless, On-column
Gas ControlElectronic Flow Control (EFC)
Temperature Programming Rate0.1 to 120 °C/min
Power Requirements230 V AC, 50/60 Hz
Detector Temperature RangeUp to 400°C (depending on detector type)

Related product manuals